Manufacturing Standards
|
|
Meets applicable safety and EMC certification standards; UL and CE certified
|
|
Arm Speed
|
|
X: >45.7 cm/sec (>18 in./sec)
Y: >40.6 cm/sec (>16 in./sec)
|
|
Communication Interface
|
|
RS-232 or GSIOC; Four inputs (contact closure, TTL, or open-collector), four relay outputs and one switched +24V DC 1A output
|
|
Injection Volume
|
|
5 µL–5 mL
|
|
Injection Reproducibility
|
|
<2.0% CV with 500-µL syringe and partial loop filling method; <5.0% CV with 500-µL syringe and total loop filling method
|
|
Liquid Handling
|
|
Reproducibility:0.57% CV with 500 µL of water
Accuracy: 99.2% with 500 µL of water
|
|
Front Panel
|
|
8-character display, START and emergency STOP soft keys
|
|
Horizontal Motion Strength
|
|
X: 5 kg (11.1 lbs.)
Y: 7 kg (15.6 lbs.)
|
|
Injection Carryover†
|
|
<0.05%
|
|
Locator Plate Capacity
|
|
Up to five Code 200-Series racks or one Code 505 or 505H rack plus one Code 200-Series rack
|
Maximum Syringe Flow Rates for Water
(with eight syringes installed)
|
|
Syringe Size
(µL)
|
Max. Flow Rate
(mL/min)
|
|
100
|
0.42
|
|
250
|
0.80
|
|
500
|
2.11
|
|
1000
|
4.22
|
|
2500
|
8.40
|
|
5000
|
12.50
|
|
Maximum Syringe Flow Rates for Water
(with four syringes installed)
|
|
Syringe Size
(µL)
|
Max. Flow Rate
(mL/min)
|
|
100
|
0.75
|
|
250
|
1.50
|
|
500
|
4.25
|
|
1000
|
8.50
|
|
2500
|
16.75
|
|
5000
|
25.00
|
|
|
Power Requirements
|
|
Frequency: 50–60 Hz
Voltage: 100–120 and 200–240V, mains voltage fluctuations not to exceed ±10% of the nominal voltage
Current rating: 2.0A for 100–120V or 1.0A for 220–240V
|
|
Probe Positioning Performance
|
|
Accuracy: ±0.5 mm in X/Y dimensions, ±1 mm in Z dimension
Repeatability: ±0.25 mm in X/Y/Z dimensions
|
|
Probe Rinse
|
|
Through a dedicated rinse station for rinsing the inside and outside of the probes
|
|
Sampler Type
|
|
X/Y/Z with stationary rack design
|
|
Software
|
|
Via Gilson UniPoint™ System Software or TRILUTION™ Software
|
|
Syringe Capacity
|
|
100, 250 or 500 µL; 1, 2.5 or 5 mL
|
|
Syringe Pump
|
|
Integral high-precision, eight-piston syringe pump
|
|
Vertical Punch Strength
|
|
0.6 kg/probe (1.4 lbs./probe)
|
|
Dimensions (w x d x h)
|
|
97.8 x 61 x 55.8* cm (38.5 x 24 x 22* in.)
*Maximum height. Z-arm height is adjustable to accommodate vessel heights between 1 and 150 mm (dependent on installed Z-arm).
|
|
Weight
|
|
39.9 kg (89 lbs.)
|
|
† Contact ManSci Customer Service to learn what methods and conditions were used to obtain the values
|