Manufacturing Standards
|
|
Meets applicable safety and EMC certification standards; UL and CE certified
|
| Bed Capacity
|
|
Twelve 96-well, 384-well |
| Communication Interface
|
|
RS-232 or GSIOC; Four inputs (contact closure, TTL, or open-collector), four relay outputs and two switched +24V DC 1A outputs
|
| Liquid Handling
|
|
Reproducibility: < 2% Accuracy: + 2% for (1-- uL-25 mL) water |
| Front Panel
|
|
Two lines of 40 alphanumeric characters, vacuum fluorescent display |
Maximum Flow Rates
|
|
Up to 25 mL/min.
(for degassed, HPLC-grade water through the probe)
|
| Power Requirements
|
|
Frequency: 50–60 Hz
Voltage: 100–120 or 220–240V, mains voltage fluctuations not to exceed ±10% of the nominal voltage
Current rating: 3.5A for 100–120V or 1.75A for 220–240V
|
| Probe Positioning Performance
|
|
Accuracy: +0.05 mm in X/Y/Z.
Repeatability: + 0.025 mm in X/Y/Z |
| Probe Rinse
|
|
Up to six solvents; optional two speed rinse pumps |
| Sampler Type
|
|
Cartesian robot featuring integrated pump, HPLC Injectors and Fraction Collection Valve |
| Software
|
|
Trilution LC Software |
| Solvent Delivery
|
|
Six position bidirectional pump , 2 uL-> 100 mL, Up to 25 mL/min. |
| Syringe Pump
|
|
Dual, internal, integral, high-precision
|
Dimensions
(w x d x h)
|
|
88.9 x 76.2 x 96.5 cm (35 x 30x 38 in.) |
| Weight
|
|
68 kg (150 lbs.) |
|
† Contact ManSci Customer Service to learn what methods and conditions were used to obtain the values.
|