215 SW Liquid Handler/Injector

Delivers the same "workhorse" performance as Gilson's 215 Liquid Handler with a smaller footprint to save on bench space, and to allow the 215 SW to fit inside a fume hood for safer use while working with volatile chemicals. The increased clearance between the Z-arm and the bed enables the use of larger tubes and bottles.

Features & Benefits
  • Accommodates virtually all of Gilson’s numerous standard and custom racks (more than 330), providing increased application flexibility


  • The 215 SW Controller provides complete remote-access control of the instrument from outside a fume hood


  • The external 402 Dual-Syringe Pump increases liquid handling and solvent delivery flexibility; large delivery volume range of 1 µL to 25 mL per piston stroke


  • Optional High-Force Z-Arm provides more than twice the vertical punch strength than the standard Z-arm for piercing thick septa


  • Optional 819 Injection Module enables automated on-line injection onto HPLC systems

845ZPREP™ System



DOWNLOAD

215 SW Liquid Handler/Injector Spec. Sheet
215 Synthesis Workstation User's Guide
Manufacturing Standards
Meets applicable safety and EMC certification standards; UL and CE certified
Arm Speed X: >45.7 cm/sec (>18 in./sec)
Y: >40.6 cm/sec (>16 in./sec)
Communication Interface
(215 SW Controller)
RS-232 or GSIOC; Four inputs (contact closure, TTL, or open-collector), four relay outputs and one switched +24V DC 1A output
Cross Contamination† 0.001% into first methanol blank
Injection Volume 1 µL–25 mL
Horizontal Motion Strength X: 4 kg (9 lbs.)
Y: 3 kg (7 lbs.)
Injection Carryover† 0.1%
Reproducibility† Liquid Handling: 0.8% at 10 µL
Injection: <0.9% CV with 1-mL syringe and total loop filling method
Locator Plate Capacity Up to four Code 200-Series racks, four Code 20- or 30-Series racks (or a combination of up to four racks of both types), or one Code 500-Series rack
Power Requirements
(215 SW Controller)
Frequency: 50–60 Hz
Voltage: 100–120 and 200–240V, mains voltage fluctuations not to exceed ±10% of the nominal voltage
Current rating: 3.0A for 100–120V or 1.5A for 220–240V
Probe Positioning Performance Accuracy: ±0.5 mm in X/Y dimensions, ±1 mm in Z dimension
Repeatability: ±0.25 mm in X/Y/Z dimensions
Probe Rinse Through a dedicated rinse station for rinsing the inside and outside of the probe
Sampler Type X/Y/Z with stationary rack design
Software Via Gilson UniPoint™ System Software or TRILUTION™ Software
Syringe Capacity 100, 250, or 500 µL; 1, 5, 10, or 25 mL
Vertical Punch Strength 5 kg (12 lbs.)
High-Force Z-Arm: 11 kg (25 lbs.)
Accuracy† Liquid Handling: 98.2% at 10 µL
Dimensions
(w x d x h)
78.7 x 61 x 74.9* cm (31 x 24 x 29.5* in.)
*Maximum height. Z-arm height is adjustable to accommodate vessel heights between 1 and 180 mm (dependent on installed Z-arm).
215 SW Controller: 26.2 x 40.6 x 15.2 cm (10.3 x 16 x 6 in.)
Weight 38.6 kg (85 lbs.)
215 SW Controller: 5.9 kg (13 lbs.)
 